Application fields
Hidex 300 SL Super Low Level
For low level applications
The Hidex 300SL is a revolutionary instrument with compact size, incorporating the most advanced triple-PMT detector technology facilitating: exceptionally high counting efficiency, luminescence free counting mode, and absolute activity counting without external radioactive source using triple-to-double coincidence ratio (TDCR) method.
Download
Product Models
Hidex 300 SL Super Low Level
For low level applications
For low level applications
- Ideal for low radioactivity applications
- Equipped with additional lead shielding and active guard
- Optimal for ³H in water, low-level environmental monitoring, radiocarbon dating & biofuel testing
- Digital Pb shield with proprietary Hidex spectral fitting algorithm
Hidex 300 SL Super Low Level
The Hidex 300 SL super low-level scintillation counter is equipped with additional lead shielding, low level PMT detectors and an active guard detector for further background reduction. The instrument is ideal for 3H in water measurements as well as other low-level environmental monitoring, radiocarbon dating and biofuel verification applications.
The advanced Active Guard is a separate scintillator detector which detects and subtracts real-time background radiation. The guard can be turned off for high energy samples and does not interfere with alpha/beta separation.
Digital Pb shield is a proprietary Hidex spectral fitting algorithm that improves counting performance by utilizing spectrum information for active sample tritium counts and background counts.
Active Guard for low level environmental measurements:
³H & 14C in natural waters
ground water dating (³H)
90Y/90Sr
gross alpha/beta
biobased 14C
low level alphas
55Fe, 66Ni
3H & 90Sr radiobioassays
Technical Specification
Sample capacity, 20mL/7mL | 40/96 |
Counting efficiency ³H/¹⁴C (%) | 70/96 |
Background ³H in water (CPM) | 3 |
Dimensions, W/H/D (cm) | 52/68/63 |
Weight (kg) | 180 |
Optional features
525-003 alpha/beta separation
462-019 External Eu-152 std source
425-2003 Cooling
Performance Specifications
All the measurements are performed at temperature of 22°C ± 2°C and at normal humidity conditions of Hidex laboratory, Turku, Finland (relative humidity not measured). Background may vary locally depending on natural environmental radiation.
Counting efficiency
Counting efficiency typical > 70% for ³H and typical > 96% for ¹⁴C with unquenched samples.
> 35% for ³H quenched (8 mL water sample + 12 mL AquaLight cocktail)
α’s (210Po, 234U/238U, 241Am, 222Rn, 226Ra) > 95 %
Typical background
3 CPM with 8 mL water + 12 mL AquaLight Low Level cocktail. *Background value measured using window with 25% counting efficiency.
< 0.3 CPM for alphas (with a/b separation option)
FOM (E2/B)
³H in water 8 + 12 ml, optimized window | > 220 |
³H in water 8 + 12 ml, with Digital Pb shield | > 300 |
³H unquenched Low Level standards, optimized window | > 350 |
¹⁴C unquenched Low Level standards, optimized window | > 1150 |
¹⁴C Benzene 3.5 ml in a mini glass vial | > 2000 |
Related methods
Read more about Hidex Methods
Industry Pioneer Radsol’s Commitment to Comprehensive Radioactivity Measurements in Korea
Hidex’s long-standing customer, Radsol, has been a pioneer in Korea’s radioactivity measurement industry, serving customers with radiochemical analysis of environmental and nuclear decommissioning samples.
Read moreRegulatory Radiation Surveillance with Hidex LSC
The National Radiation Protection Institute (NRPI), Prague continues to trust Hidex’s radiation detection equipment for needs of the Czech regulatory body (SÚJB) and for research studies.
Read more
1 / 2